Accu-Chek Insight product support

The Accu-Chek Insight system is designed for easy and discreet insulin pump therapy.

More product information

How to use the Accu-Chek Insight insulin pump system

To learn more about the basics of insulin pump therapy and how to use the Accu-Chek Insight system visit our online training tool.

Accu-Chek Insight diabetes therapy system web-training

For further information on how to use your Accu-Chek Insight system, please see the owner's booklets and related training materials for your Accu-Chek Insight handset and Accu-Chek Insight insulin pump system.

Error codes

E6 – Mechanical Error

The mechanical system of the pump does not work properly. Change the battery and the cartridge.

E7 – Electronic Error

When this message occurs, change the battery. If the message still occurs, you cannot resolve the problem yourself. Contact the Accu-Chek Pump Careline for advice. Change to an alternative insulin therapy.

M24 – Occlusion

The insulin cannot flow freely. Change the complete infusion set. If the Maintenance message occurs again, change the cartridge as well.

M27 – Communication lost

The meter and pump have timed out while waiting for pairing confirmation. Repeat the steps from the Pairing the Meter and Pump section of your user manual or download the pairing instructions here, and follow the on-screen instructions on both devices.

M68 – Connection lost

One of the following has occurred:
  • The pump battery has died
  • The pump and handset are out of communication range
  • A pump key has been pressed, temporarily disabling communication between the handset and pump 

You may need to:
  • Change the pump battery
  • Ensure that the handset and pump are no further than 2 metres apart
  • Wait for the pump display to timeout and the handset and pump to automatically resume communication

M83 – Pump not able to complete task

Communication has been lost between the handset and pump. When this message is displayed during an attempted bolus insulin delivery by the pump, perform the following:
  1. Press “OK” on the M83 message to confirm it
  2. Power down the handset
  3. Check the pump to determine the amount of insulin that was actually delivered by the pump
  4. Re-establish communication between the handset and pump by waiting for the pump display to timeout and then power up the handset. If the handset does not connect to the pump, the handset can be reset by holding the power button for 5 seconds
  5. Make sure you do not program a duplicate bolus. If the original bolus amount was not fully delivered by the pump, program the undelivered bolus amount 

E57 – Electronic error

A temporary error has been identified
  • Please ensure that you insert the test strip correctly. The handset will then check that the test strip is suitable to use for testing. Do not apply your drop of blood on the test strip before the “Apply Drop” screen appears. Do not move the test strip during testing
  • Often, resetting the handset by holding the on/off button for 5 seconds or more can stop the E57 error
  • If frequent errors occur, please contact the Accu-Chek Pump Careline 

Frequently asked questions

Can I disconnect from my insulin pump temporarily?

Because Accu-Chek insulin pumps can be disconnected at the infusion site you can remove the pump temporarily (for up to an hour) to bathe, go swimming or take part in contact sports. 
Always remember to check your blood glucose and correct as needed before disconnecting your insulin pump and again when you reconnect. If you are intending to be disconnected from your pump for longer than an hour, you should check your blood glucose regularly throughout the time you are disconnected, reconnecting and correcting as necessary. 
You should discuss your personal requirements for disconnecting from your insulin pump with your healthcare professional before going ahead.

Can I swim, shower or bathe while wearing an insulin pump?

Both the Accu-Chek Combo and Accu-Chek Insight insulin pumps have an IPX8 rating. This means that both pumps are protected from brief immersion in water. As small hairline cracks in the housing of the device, or adapters and battery caps not being tightened enough could cause water to get into the device and cause damage, we recommend that you disconnect your insulin pump and leave it in a safe place before swimming, showering or bathing. This should only be done following advice from your healthcare professional, and you should test and correct as necessary at least every hour.

What supplies should I take with me when I go on holiday?

When going on holiday, you should always take extra supplies to make sure that you are covered for anything that might happen during your trip. Remember to take the following with you when you go away:
  • Plenty of cannula and infusion sets
  • Spare batteries
  • Insulin and pump cartridges
  • Test strips and lancets
  • Emergency contact details
  • Fast acting carbohydrate to treat hypos
  • Travel letter from hospital / diabetes team
  • Details of how to manage illness / pump failure or loss
  • Back up insulin pens, cartridges and needles
download the Accu-Chek Travel guide

Can MRI or X-ray machines disrupt the system?

Some types of medical examinations, such as MRIs, CT scans and X-Rays, may interfere with the operation of your insulin pump. As such, you should disconnect the insulin pump from the body and leave it in a safe place outside of the treatment room. 
If you use a steel infusion set, you may also be required to remove the infusion set for some procedures. Please consult with a healthcare professional prior to the procedure.  

I noticed my cartridge and tubing have air bubbles in them. What could be the cause?

Bubbles naturally occur in all liquids, including insulin, when they undergo a fairly rapid temperature change, i.e. warm to cool. To help to prevent bubbles, remove insulin from the fridge at least an hour before use. If you are self-filling cartridges, make sure to fill the cartridge with room-temperature insulin and draw it from the vial slowly and steadily. You should regularly check your cartridge and tubing for bubbles and prime out any that you find. Always remember to disconnect from your insulin pump before priming. 

How long is my pump warranty?

The technical warranty of your Accu-Chek insulin pump is 4 years from the day that you start on your insulin pump. It covers technical issues with the device, but does not cover loss or damage due to misuse.

Can I wear my insulin pump during sports?

Your insulin pump allows you to quickly and precisely adjust your insulin levels before, during and after exercise, to carefully match your individual needs. 
All Accu-Chek insulin pumps are suitable to use during most types of exercise, however we recommend that you remove the insulin pump before taking part in any contact sports (e.g. rugby or martial arts) in order to prevent damage to the device or harm to yourself. 

What do I need to think about when going to a theme park with my pump?

Many theme park rides, such as rollercoasters, are operated by high powered magnetic equipment, which have the potential to affect your insulin pump. Please check with the Theme Park Occupational Health and Safety officer about the magnetic equipment they use and what distance in meters / feet the field surrounds them. When within these fields do not wear or operate your insulin pump.
Please disconnect from your insulin pump and leave it in a safe place until you are no longer within the range of these magnets. If there is nowhere safe for you to leave your insulin pump, you should take it with you, however you will need to put it in STOP mode and disconnect it from your body before entering the field of exposure. Leave your insulin pump in STOP mode until you are outside the magnetic field. When it is safe to do so, please reconnect and START your insulin pump. Please remember to check your blood glucose levels before and after entering the field of exposure.
Always carry backup supplies – including hyper / hypo treatments and a backup insulin pen with you. If you are disconnected from your insulin pump for longer than 30 – 60 minutes you may need to reconnect to your insulin pump and give a correction dose (if safe to do so) or give a correction dose with your back up insulin delivery mechanism if you are still within the exposure range. You should follow your healthcare professional’s instructions to replace missed insulin or to correct a hypo.

Can I go in a sauna or steam room while wearing my pump?

No. Due to the high temperatures in saunas and steam rooms we recommend that you disconnect your pump and also remove your infusion set before entering. High temperatures can affect the insulin within your pump and also potentially cause damage to the pump itself.
Please consult with your healthcare professional to determine what precautions you should take prior to entering a sauna or steam room.