Accu-Chek Performa Nano
Error codes E-10
The time and date settings may be incorrect. Make sure the time and date are correct and adjust if necessary.
The time and date settings may be incorrect. Make sure the time and date are correct and adjust if necessary.
The batteries are almost out of power. Change the batteries now. If the message reappears after the batteries have been replaced, slide the battery drawer out of the meter, press any meter button, then reinsert the battery drawer back into position.
The Accu-Chek TenderLink infusion set is compatible with the Accu-Chek Combo Insulin Pump.
Check other FAQ categories by clicking on the below category names below
The temperature is above or below the proper range for the system. Refer to the test strip package insert for system operating conditions. Move to an area with the appropriate conditions, wait 5 minutes, and repeat the blood glucose test or control test. Do not artificially heat or cool the meter.
Please watch the video on this page and refer to the Accu-Chek TenderLink quick reference guide for the steps required to insert the Accu-Chek TenderLink infusion set.
Check other FAQ categories by clicking on the below category names below
An electronic error occurred, or in rare cases, a used test strip was removed and reinserted. Turn the meter off and on, or take the battery out for 20 seconds and reinsert it. Perform a blood glucose or control test.
Blood or control solution was applied to the test strip before the flashing drop symbol appeared on the display. Discard the test strip and repeat the blood glucose or control test.
This message may appear when using a white code chip in the meter. It means the white code chip is from an expired lot of test strips. Turn the meter off and insert a black activation chip. Make sure the time and date in the meter are correct.
Not enough blood or control solution was drawn into the test strip for measurement or was applied after the test had started. Discard the test strip and repeat the blood glucose or control test.
The user's blood glucose may be extremely high or a meter or a test strip error has occurred.